The Press

Hospital steriliser was never on

- Marty Sharpe marty.sharpe@stuff.co.nz

A broken switch, a broken printer and a raft of human and systems errors were to blame for inadequate­ly sterilised surgical equipment being used on patients in Hawke’s Bay, reports have found.

The region’s district health board yesterday released internal and external reports on the incident that saw the equipment used on up to 55 patients over nine days in February.

The reports found that staff who put the equipment in one of four sterilisin­g machines, or autoclaves, at about 10.45pm on February 1 did not turn the machine on.

Either way, the autoclave makes a loud beeping noise when it starts and staff should have ensured this occurred. They could not recall if they heard it.

The autoclaves are all connected to their own printers which signalled that the machines had been started and that they had been completed.

The printer on the machine in question had been broken for several months so no signal was given and no print out was available at completion. Instead, staff had to check on a computer to see if the machine had been operating.

On the morning of February 2 a different staff member removed the equipment from the autoclave, without adequately checking if it had been working.

The staff member did not check the computer to see if the equipment had been sterilised before dispatchin­g the equipment for use in theatres.

The staff member also failed to check sterilisat­ion codes on packages had not changed colour, indicating they had been sterilised.

Then clinical staff who were taking the equipment from their packs failed to check the colour codes before use.

It wasn’t until February 11 that a nurse raised the alarm. The external review found that all three sterilisat­ion checks failed and that staff failed to follow sterilisat­ion policy.

The start button and printer have been fixed and several changes to systems have been put in place. The review found that sterile services nationwide had been ignored for ‘‘too long’’ and the Health Ministry should immediatel­y put in place changes, including electronic tracking of each individual instrument and external auditing of all DHB sterilisat­ion units.

The DHB’s executive director provider services Colin Hutchison said the review made it clear that no one person or department could be held accountabl­e as there were ‘‘many errors across a number of systems and processes’’.

None of the affected patients have shown infection signs, and the autoclave was part of a four stage sterilisat­ion so the chance of infection was negligible.

 ??  ?? One of the four steriliser­s, or autoclaves, at Hawke’s Bay Hospital was not turned on on the night of February 1, a review has found. DHB executive director provider services Colin Hutchison cited many errors.
One of the four steriliser­s, or autoclaves, at Hawke’s Bay Hospital was not turned on on the night of February 1, a review has found. DHB executive director provider services Colin Hutchison cited many errors.